Detection of dimethylamine in the low pptv range using nitrate chemical ionization atmospheric pressure interface time-of-flight (CI-APi-TOF) mass spectrometry

Show simple item record

dc.contributor.author Simon, Mario
dc.contributor.author Heinritzi, Martin
dc.contributor.author Herzog, Stephan
dc.contributor.author Leiminger, Markus
dc.contributor.author Bianchi, Federico
dc.contributor.author Praplan, Arnaud
dc.contributor.author Dommen, Josef
dc.contributor.author Curtius, Joachim
dc.contributor.author Kuerten, Andreas
dc.date.accessioned 2016-08-25T14:25:01Z
dc.date.available 2016-08-25T14:25:01Z
dc.date.issued 2016
dc.identifier.citation Simon , M , Heinritzi , M , Herzog , S , Leiminger , M , Bianchi , F , Praplan , A , Dommen , J , Curtius , J & Kuerten , A 2016 , ' Detection of dimethylamine in the low pptv range using nitrate chemical ionization atmospheric pressure interface time-of-flight (CI-APi-TOF) mass spectrometry ' , Atmospheric Measurement Techniques , vol. 9 , no. 5 , pp. 2135-2145 . https://doi.org/10.5194/amt-9-2135-2016
dc.identifier.other PURE: 67750924
dc.identifier.other PURE UUID: 42da1cae-6284-4de5-aff4-d13f36aa946f
dc.identifier.other WOS: 000377855300012
dc.identifier.other Scopus: 84970027715
dc.identifier.other ORCID: /0000-0003-2996-3604/work/31537484
dc.identifier.uri http://hdl.handle.net/10138/166164
dc.description.abstract Amines are potentially important for atmospheric new particle formation, but their concentrations are usually low with typical mixing ratios in the pptv range or even smaller. Therefore, the demand for highly sensitive gas-phase amine measurements has emerged in the last several years. Nitrate chemical ionization mass spectrometry (CIMS) is routinely used for the measurement of gas-phase sulfuric acid in the sub-pptv range. Furthermore, extremely low volatile organic compounds (ELVOCs) can be detected with a nitrate CIMS. In this study we demonstrate that a nitrate CIMS can also be used for the sensitive measurement of dimethylamine (DMA, (CH3)(2)NH) using the NO3-center dot(HNO3)(1-2)center dot(DMA) cluster ion signal. Calibration measurements were made at the CLOUD chamber during two different measurement campaigns. Good linearity between 0 and similar to 120 pptv of DMA as well as a sub-pptv detection limit of 0.7 pptv for a 10 min integration time are demonstrated at 278 K and 38% RH. en
dc.format.extent 11
dc.language.iso eng
dc.relation.ispartof Atmospheric Measurement Techniques
dc.rights cc_by
dc.rights.uri info:eu-repo/semantics/openAccess
dc.subject PARTICLE FORMATION EVENTS
dc.subject SULFURIC ACID-AMINE
dc.subject H2SO4 PRODUCTION
dc.subject BOREAL FOREST
dc.subject NUCLEATION
dc.subject AEROSOL
dc.subject AMMONIA
dc.subject CLUSTERS
dc.subject TRIMETHYLAMINE
dc.subject RATES
dc.subject 114 Physical sciences
dc.title Detection of dimethylamine in the low pptv range using nitrate chemical ionization atmospheric pressure interface time-of-flight (CI-APi-TOF) mass spectrometry en
dc.type Article
dc.contributor.organization Helsinki Institute of Physics
dc.description.reviewstatus Peer reviewed
dc.relation.doi https://doi.org/10.5194/amt-9-2135-2016
dc.relation.issn 1867-1381
dc.rights.accesslevel openAccess
dc.type.version publishedVersion

Files in this item

Total number of downloads: Loading...

Files Size Format View
amt_9_2135_2016.pdf 531.6Kb PDF View/Open

This item appears in the following Collection(s)