Direct tomography imaging for inelastic x-ray scattering experiments at high pressure

Show full item record



Permalink

http://hdl.handle.net/10138/175015

Citation

Sahle , C J , Rosa , A , Rossi , M , Cerantola , V , Spiekermann , G , Petitgirard , S , Jacobs , J , Huotari , S , Moretti Sala , M & Mirone , A 2017 , ' Direct tomography imaging for inelastic x-ray scattering experiments at high pressure ' , Journal of Synchrotron Radiation , vol. 24 , no. 1 , pp. 269-275 . https://doi.org/10.1107/S1600577516017100

Title: Direct tomography imaging for inelastic x-ray scattering experiments at high pressure
Author: Sahle, Christoph Johannes; Rosa, Angelica; Rossi, Matteo; Cerantola, Valerio; Spiekermann, Georg; Petitgirard, Sylvain; Jacobs, Jeroen; Huotari, Simo; Moretti Sala, Marco; Mirone, Alessandro
Contributor: University of Helsinki, ESRF - European Synchrotron Radiation Facility
University of Helsinki, Department of Physics
Date: 2017-01-01
Language: eng
Number of pages: 7
Belongs to series: Journal of Synchrotron Radiation
ISSN: 0909-0495
URI: http://hdl.handle.net/10138/175015
Abstract: A method to separate the non-resonant inelastic X-ray scattering signal of a micro-metric sample contained inside a diamond anvil cell (DAC) from the signal originating from the high-pressure sample environment is described. Especially for high-pressure experiments, the parasitic signal originating from the diamond anvils, the gasket and/or the pressure medium can easily obscure the sample signal or even render the experiment impossible. Another severe complication for high-pressure non-resonant inelastic X-ray measurements, such as X-ray Raman scattering spectroscopy, can be the proximity of the desired sample edge energy to an absorption edge energy of elements constituting the DAC. It is shown that recording the scattered signal in a spatially resolved manner allows these problems to be overcome by separating the sample signal from the spurious scattering of the DAC without constraints on the solid angle of detection. Furthermore, simple machine learning algorithms facilitate finding the corresponding detector pixels that record the sample signal. The outlined experimental technique and data analysis approach are demonstrated by presenting spectra of the Si L-2,L-3-edge and O K-edge of compressed alpha-quartz. The spectra are of unprecedented quality and both the O K-edge and the Si L-2,L-3-edge clearly show the existence of a pressure-induced phase transition between 10 and 24 GPa.
Subject: 114 Physical sciences
Rights:


Files in this item

Total number of downloads: Loading...

Files Size Format View
jsy2hf5329.pdf 729.9Kb PDF View/Open

This item appears in the following Collection(s)