An Oral Rinse Active Matrix Metalloproteinase-8 Point-of-Care Immunotest May Be Less Accurate in Patients with Crohn’s Disease

Näytä kaikki kuvailutiedot



Pysyväisosoite

http://hdl.handle.net/10138/313296

Lähdeviite

Rautava, J.; Gürsoy, U.K.; Kullström, A.; Könönen, E.; Sorsa, T.; Tervahartiala, T.; Gürsoy, M. An Oral Rinse Active Matrix Metalloproteinase-8 Point-of-Care Immunotest May Be Less Accurate in Patients with Crohn’s Disease. Biomolecules 2020, 10, 395.

Julkaisun nimi: An Oral Rinse Active Matrix Metalloproteinase-8 Point-of-Care Immunotest May Be Less Accurate in Patients with Crohn’s Disease
Tekijä: Rautava, Jaana; Gürsoy, Ulvi K.; Kullström, Adrian; Könönen, Eija; Sorsa, Timo; Tervahartiala, Taina; Gürsoy, Mervi
Julkaisija: Multidisciplinary Digital Publishing Institute
Päiväys: 2020-03-04
URI: http://hdl.handle.net/10138/313296
Tiivistelmä: The diagnostic accuracy of point-of-care (PoC) applications may be compromised in individuals with additional inflammatory conditions. This cross-sectional study examined the performance of a commercial oral rinse active matrix metalloproteinase-8 (aMMP-8) PoC immunotest in individuals with (<em>n</em> = 47) and without Crohn&rsquo;s disease (CD) (<em>n</em> = 41). Oral rinse collected from the participants was analyzed by the PoC immunotest. Molecular forms and fragments of salivary MMP-8 were detected by western immunoblotting. The sensitivity of the immunotest for periodontitis was 60.0% in the CD group and 90.0% in the control group. The respective specificity was 75.0% and 80.0%. In both groups, clinical diagnosis of periodontitis exhibited a significant association with the immunotest results, however, the odds ratio (OR) was more than ten-fold in controls (OR 54.3, 95% CI: 3.1&ndash;953, <em>p </em>= 0.006) in comparison to CD patients (OR 5.2, 95% CI: 1.3&ndash;21.6, <em>p </em>= 0.022). According to Western immunoblot results, the immunotest MMP-8 positivity was not related to elevated levels of molecular forms and fragments of MMP-8 in the CD group, as in the control group. The diagnostic accuracy of the aMMP-8 PoC oral rinse immunotest is reduced in CD patients, which may be related to lower levels or undetectable complexes.


Tiedostot

Latausmäärä yhteensä: Ladataan...

Tiedosto(t) Koko Formaatti Näytä
biomolecules-10-00395.pdf 1.431MB PDF Avaa tiedosto

Viite kuuluu kokoelmiin:

Näytä kaikki kuvailutiedot