Tunninen , V , Kauppinen , T & Eskola , H 2018 , Physical characteristics of collimators for dual-isotope imaging with Tc-99m and I-123 . in H Eskola , O Väisänen , J Viik & J Hyttinen (eds) , EMBEC & NBC 2017 . IFMBE Proceedings , vol. 65 , SPRINGER-VERLAG SINGAPORE PTE LTD , Singapore , pp. 245-249 , Joint Conference of the European Medical and Biological Engineering Conference (EMBEC) / Nordic-Baltic Conference on Biomedical Engineering and Medical Physics (NBC) , Tampere , Finland , 01/06/2017 . https://doi.org/10.1007/978-981-10-5122-7_62
Title: | Physical characteristics of collimators for dual-isotope imaging with Tc-99m and I-123 |
Author: | Tunninen, Virpi; Kauppinen, Tomi; Eskola, Hannu |
Other contributor: |
Eskola, Hannu
Väisänen, Outi Viik, Jari Hyttinen, Jari |
Contributor organization: | Department of Diagnostics and Therapeutics Clinicum HUS Medical Imaging Center |
Publisher: | SPRINGER-VERLAG SINGAPORE PTE LTD |
Date: | 2018 |
Language: | eng |
Number of pages: | 5 |
Belongs to series: | EMBEC & NBC 2017 |
Belongs to series: | IFMBE Proceedings |
ISBN: | 978-981-10-5121-0 978-981-10-5122-7 |
ISSN: | 1680-0737 |
DOI: | https://doi.org/10.1007/978-981-10-5122-7_62 |
URI: | http://hdl.handle.net/10138/327120 |
Abstract: | The purpose of this study was to compare the physical characteristics of Low Energy High Resolution (LEHR), Low Energy Ultra High Resolution (LEUHR) and Medium Energy Low Penetration (MELP) collimators for simultaneous Tc-99m and I-123 imaging. MELP collimator performed well with I-123 high-energy gamma photons, but low resolution makes it unsuitable to use for acquisition of small structures such as parathyroid adenomas. LEUHR collimators optimized for Tc-99m have highest resolution, but the differences in septal penetration and sensitivity in favor of LEHR collimator needs to be tested with specific parathyroid phantoms. |
Subject: |
dual-isotope
Tc-99m I-123 collimator SPECT/CT MEDIUM-ENERGY COLLIMATORS SUBTRACTION SPECT HYPERPARATHYROIDISM PHANTOM 217 Medical engineering 3111 Biomedicine |
Peer reviewed: | Yes |
Usage restriction: | openAccess |
Self-archived version: | acceptedVersion |
Total number of downloads: Loading...
Files | Size | Format | View |
---|---|---|---|
Tunninen_Physical_properties.pdf | 372.7Kb |
View/ |